Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–23
Page 4 of 20 (Imaging Parameter 4 – Probe Dependent 4) (cont’d)
Auto Optimize with Auto Angle Correct EC, UP, R, P
Choose the default setting for the Auto Angle correction while ASO is active.
On or Off.
Auto Optimize with Doppler Auto Invert EC, UP, R, P
Choose the default setting for the Doppler Auto Invert function while ASO is active.
On or Off.
Page 5 of 20 (Imaging Parameter 5 – Probe Dependent 5)
CWD Gain EC, UP, P, R
Choose the default gain value for continuous wave Doppler Mode.
Value from 0 to 32 in 2 digit increments.
CWD Dynamic Range EC, UP, P, R
Choose the continuous wave Doppler Mode dynamic range default value.
Value from 18 to 48 in 6 digit increments.
CWD Gray Scale Map EC, UP, P, R
Choose the continuous wave Doppler Mode gray scale default map.
1, 2, 3, 4, 5, 6, 7, 8, 9 or 10.
CWD Rejection EC, UP, P, R
Choose the continuous wave Doppler Mode rejection default level.
Level from 0 to 40 in 2 digit increments.
CWD Velocity [cm/s] EC, UP, P, R
Choose the continuous wave Doppler Mode velocity scale default value.
Value from 10 to 1000 in 10 digit increments.
CWD Wall Filter [cm/s] EC, UP, P, R
Choose the continuous wave Doppler Mode wall filter default value.
0.3, 0.5, 0.7, 1.0, 1.5, 2.0, 3.0, 5.0, 7.0, 10, 15, 20, 30 or 50.