Site Requirements
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
3–4
Before the system arrives
NOTICE
This medical equipment is approved, in terms of the prevention
of radio wave interference, to be used in hospitals, clinics and
other institutions which are environmentally qualified. The use
of this equipment in an inappropriate environment may cause
some electronic interference to radios and televisions around
the equipment. This equipment can be used in residential
areas only under the supervision of physicians or qualified
technicians.
Ensure that the following is provided for the new system:
A separate power outlet with a 20 amp circuit breaker for
120 VAC (USA) or 7.5 amp circuit breaker for 220–240 VAC
(Europe, Latin America).
Take precautions to ensure that the console is protected
from electromagnetic interference.
Precautions include:
Operate the console at least 15 feet away from motors,
typewriters, elevators, and other sources of strong
electromagnetic radiation.
Operation in an enclosed area (wood, plaster or
concrete walls, floors and ceilings) helps prevent
electromagnetic interference.
Special shielding may be required if the console is to be
operated in the vicinity of radio broadcast equipment.