Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–22
Page 4 of 20 (Imaging Parameter 4 – Probe Dependent 4) (cont’d)
PWD Alternative Scan EC, UP, P
Choose the action of the B-Pause key while in PWD Mode.
On
Pressing B-Pause toggles between B-Mode frozen/PWD real-time or B-Mode
real-time/PWD frozen.
Off
Pressing B-Pause freezes or unfreezes the B-Mode image only.
PWD/CWD Flow Angle [deg] EC, UP, P, R
Choose the default for the Doppler flow
θ
angle.
Value from -80 to +80 in 5
°
increments.
CFM/PWD Ratio in Triplex EC, UP, P, R
Choose the velocity ratio between the CFM display and Pulsed Doppler.
1/1
Same velocity display.
1/2
Pulsed Doppler velocity 2 times CFM.
1/4
Pulsed Doppler velocity 4 times CFM.
PDI/PWD Ratio in Triplex EC, UP, P, R
Choose the velocity ratio between the PDI display and Pulsed Doppler.
1/1
Same velocity display.
1/2
Pulsed Doppler velocity 2 times PDI.
1/4
Pulsed Doppler velocity 4 times PDI.
CFM/PDI Shrink in Triplex EC, UP, P, R
Choose to turn CFM/PDI shrink function on or off. Allows for shrinking of the CFM/PDI
window to a specified value to improve frame rate.
On or Off.
CFM/PDI Shrink Angle/Width [deg, mm] EC, UP, P, R
Choose the shrink value for the CFM/PDI window in degrees for convex and sector
probes or millimeters for linear probes.
Value range depends on probe type.
CFM/PDI Performance Priority EC, UP, R, P
Used to avoid the color sensitivity deterioration in Triplex mode for the MR–Flow
supported probes.
Max Res.
Best resolution and sensitivity (Version 5.00y level).
Triplex_Sens.
Best color sensitivity in Triplex mode for the MR–Flow supported
probes.
NOTE: The color sensitivity does not change for the M12L and C358 probes, even if the
“Triplex_Sens.” is selected.