Errors
Error Table
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-22
September 2016
Error Table
Table 10.6
Displayed or Printed Error Messages
Error #
Abbreviation/
Message
Explanation
Action
E1
No AGXS Data
No Antigen Excess Data
Points
CALL BECKMAN COULTER
CLINICAL SUPPORT
E2
No AGXS Data
No Antigen Excess Data
Points High
CALL BECKMAN COULTER
CLINICAL SUPPORT
E3
Motion Error
Motion Error
1. Remove any obstructions from
moving parts.
2. Home system.
3. Follow Event Log directions to
display the Motion Error Event
Log.
4. Restart the run.
Report any failed devices to Beckman
Coulter Clinical Support.
E4
No Dil Wells
Out of Dilution Wells
From Status menu, check status of
dilution segments and replace
segments showing zero ("0") status.
E5
RGT Cart N/A
Insufficient Chemistry
Reagent and/or
Uncalibrated Reagent
If reagent volume is low, load a new
reagent cartridge, and Read Reagents
from the Rgts/Cal screen.
If reagent is uncalibrated, calibrate.
E6
Back to Back
Excessive Calibrator
Replicate Spread
Repeat calibration.
E7
Scale Fct
Scale Factor Range Error Repeat calibration.
E8
Unstable RXN
Unstable Reaction
CALL BECKMAN COULTER
CLINICAL SUPPORT
E9
RGT Cart N/A
Insufficient Chemistry
Reagent
Reagent not on board. Load a new
reagent cartridge, and Read Reagents
from the Rgts/Cal screen.
E10
Chk Optics
Optics Error for AGXS,
excessive Q Values
CALL BECKMAN COULTER
CLINICAL SUPPORT
10
Code Segment
CRC Failure
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
(1 of 21)