Defining a Control
Entering the Minimum Control Definition
Quality Control
IMMAGE 800 Operations Manual A11403
Page 9-8
September 2016
11
Enter the Mean value beside the chemistry it corresponds to in the
Assigned Mean
field. Press
[Enter]
.
NOTICE
The Assigned Mean is set to "0" and the Assigned SD is set to "99999" if
"0" is entered in the Assigned Mean field.
12
Enter the "1" SD value beside the chemistry it corresponds to in the
Assigned SD
field. Press
[Enter]
.
13
Proceed to Defining More Information on a Control to continue adding
information to the control definition.
OR
Select any icon from the menu bar to exit the Define/Edit Controls screen
and save the control definition.
14
A message will display upon exiting the screen if there is any missing
information for the control definition. (Refer to Figure 9.4.)
Select
<OK>
to exit without saving the control.
OR
Select
<Cancel>
to return to the Define/Edit screen to add the necessary
information.
15
A Define/Edit Controls dialog box is displayed when:
• the
Assigned Mean
field and/or
Assigned SD
field are set to "0", or
• there are no values entered in the
Assigned Mean
field and/or
Assigned
SD
field.
Select…
to…
<1> Save file(s) with
Mean = 0.0, SD = 99999
set the mean to "0" and the SD to
"99999" for the chemistries in
question.
<2> Delete undefined file(s)
from this control
delete the chemistries in question
from the control.
<Cancel>
to return to the Define/Edit screen
and enter the mean and/or SD.
Step
Action
, continued
(2 of 2)