Programming Rate Mode
Introduction
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-92
September 2016
Programming Rate Mode
Introduction
After the UDR chemistry protocol is defined and the chemistry is configured, Rate
Mode is used to optimize UDR parameters. After running the UDR chemistry, a
report with only instrument responses (IR) is automatically generated. A maximum of
six UDRs may be programmed at one time in rate mode. Rate Mode can be run on a
calibrated or uncalibrated UDR.
Programming Rate Mode
The instrument status must be in
Standby
in order to proceed with the steps below to
program a UDR for rate mode.
Step
Action
1
From the User-Defined Chemistry screen, select
Rate Mode [F3]
. (Refer
2
Select up to six of the UDR chemistries.
3
Select
<OK>
to continue. Rate Mode may be run on a calibrated or
uncalibrated UDR.
NOTICE
Rate results are always uncalibrated rate, whether the UDR is calibrated
or uncalibrated.
OR
Select
<Cancel>
to exit rate mode and return to the User-Defined
Chemistry screen.
4
From the UDR Rate Mode Assign screen (Refer to Figure 5.40.), enter an
available
rack number in the
Rack
field.
OR
Select
Clear Racks [F1]
and Enter the racks and/or positions to clear.
Select
<OK>
to clear or
<Cancel>
to exit without clearing.
5
Enter a position number in the
Pos
field. Up to nine positions may be
entered.
6
Enter a Sample ID in the
Sample ID
field for each position assigned. Up
to 15 characters may be entered.
7
Select
Save [F9]
to save the program.
OR
Select
Cancel [F10]
to exit the screen without saving the program.
8
Repeat Steps 1-7 to program additional racks for rate mode.
9
Print a Load List from Sample Programming.
10
Load the samples on the sample carousel.
(1 of 2)