Bar Code Setup
Restoring Defaults
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-16
September 2016
Restoring Defaults
The instrument status must be in
Standby
in order to proceed with the steps below to
restore bar code symbology defaults.
Additional Information
CHAPTER 7,
Sample Programming
, Selecting Save/Next, Programming a Batch of
Samples, Identifying a Batch
Step
Action
1
From the Setup screen, select
<3> Bar Code
.
2
Select
Restore Default [F1]
to return all bar code setups to their defaults.
3
Select
<OK>
at the confirmation screen to restore the defaults.
OR
Select
<Cancel>
to return to the Bar Code Setup screen without restoring
the defaults.