As-Indicated Maintenance
As-Indicated
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-14
September 2016
As-Indicated Maintenance
As-Indicated
The following table lists the As-Indicated maintenance for the IMMAGE 800 System.
Decontaminating Carousels and Racks
To decontaminate the sample or reagent carousels or sample racks, perform the
following steps. Verify from the status bar that the instrument status is in
Standby
.
Table 10.4
As-Indicated Maintenance
Clean
Perform
Clean or replace printer cartridge
(Refer to Printer Manual.)
Replace all Cuvettes every 10,000 tests.
For systems using DIL2, cuvettes may need to be
replaced more frequently. (Refer to IMMAGE
Immunochemistry Systems,
Chemistry
Information Manual
, Appendix F,
System Reagent
Configuration and Part Numbers
for chemistries
using DIL2.)
Decontaminate Sample/Reagent
Carousels and Sample Racks
Cuvette Wash.
Replace syringes if leaking occurs, or as needed.
Replace UPS battery when "battery low" light or
audible alarm is activated. (Refer to UPS User’s
Manual.)
NOTICE
DO NOT Autoclave Carousels or Sample Racks