Overview
Introduction
IMMAGE 800 Operations Manual A11403
Sample Programming
September 2016
Page 7-79
7
Routine Operation
Overview
Introduction
The following tables summarize routine operating procedures for various bar code and
host computer capabilities. Bar Code Setup and Host Communications Setup should
first be appropriately defined. (Refer to CHAPTER 5,
Operating in Host Query Mode
Follow the steps below for routine operation of the IMMAGE 800 Immunochemistry
System using host query.
Step
Action
1
Check and clear as necessary:
• sample racks
Check, clear, and replace as necessary:
• dilution segments
Check:
• reagents, buffers, and diluents
• wash solution volume
(Refer to Preparing for Programming/Running in this chapter.)
2
Request calibrations if necessary. (Refer to CHAPTER 6,
.)
(1 of 2)