Defining a UDR Chemistry
Description of Definition Fields
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-74
September 2016
Defining a UDR Chemistry
Description of Definition Fields
The following table describes the fields of the Define/Edit User-Defined Chemistry
screen, Page 1. (Refer to Figure 5.32.) Use the tab key to navigate.
Table 5.7
Protocol Definition Fields, Page 1
Field
Entries Allowed
Function
Chem Name
Two to five alphanumeric
characters
Unique name for chemistry.
Reagent Lot Number
A maximum of eight
alphanumeric characters
Identifies reagent lot number.
Cartridge Lot Number
A maximum of eight
alphanumeric characters
Identifies cartridge lot number.
Cartridge Serial
Number
A maximum of four
alphanumeric characters
Identifies unique serial number
of UDR cartridge.
AGXS Limit
1-9999
Identifies Antigen Excess limit
rate.
Comments
A maximum of 20
alphanumeric characters
Provides additional vendor
information, or other comments.
Units
Selection from list
Concentration Units of UDR
results.
Conversion Factor
Not applicable
• Displays the conversion
factor input from the Units
option button.
Protocol
Selection from list:
• Non-Competitive
nephelometric
• Competitive
nephelometric
• Non-Competitive
NIPIA
• Competitive NIPIA
• Immunoprecipitin reaction
detected by rate
nephelometry.
• Inhibition immunoprecipitin
reaction detected by rate
nephelometry.
• Immunoprecipitin reaction
detected by rate turbidimetry.
• Inhibition immunoprecipitin
reaction detected by rate
turbidimetry.
Reagent Expiration
Date
Date in order as defined
in Date Setup
Reagent will be flagged on
reports as expired after this date.
(1 of 2)