Approving a Calibration
Approving a UDR Calibration
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-107
5
Figure 5.53
User-Defined Model Dialog Box
Step
Action
1
From the User-Defined Chemistries screen, select
Approve Cal [F5]
.
2
Use
<Page Up>
or
<Page Down>
to review the plots of the models for the
calibration. The last page has a composite plot of all the models.
3
To print a curve-fit model plot, press
[Ctrl, P]
on the keyboard.
4
Select
Model [F1]
to display the curve-fit model options. (Refer to Figure
5
Select the number beside the model desired and select
<OK>
to approve
the UDR calibration.
OR
Select
<Cancel>
to exit the User-Defined Model dialog box.
6
Select
Print Report [F8]
to print the data and statistics for a curve-fit
model.
7
Program controls on the UDR through Sample Programming to see if they
are acceptable with the chosen calibration model before clearing the UDR
cal rack. To test another model with controls, repeat Steps 1-7.
8
When the final calibration model has been selected, clear the UDR
calibration rack. Refer to "Clearing a UDR Cal Rack" in this chapter.