Loading Wash Solution
Introduction
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-13
6
Loading Wash Solution
Introduction
The wash solution is used for rinsing the probes, mixers and cuvettes on the IMMAGE
800.
When to Load
Load a new box of wash solution when the volume of the solution is low
(approximately 1000 tests/10 liter box).
Loading the Wash Solution
The instrument status must be in
Standby
in order to proceed with the steps below to
load the wash solution.
Step
Action
1
Remove the screw cap, with the tubings and straw attached, from the wash
solution container in use.
2
Remove the screw cap from the new container of wash solution.
3
Place the cap with the tubings and straw attached into the new wash
solution and screw on the cap. The blue tubing is attached to the side with
a straw, the orange tubing is not attached to a straw.
4
Verify that the blue and orange tubings are attached to the top of the cap.
5
No software intervention is required. Priming is not required.