Checking Calibration Status
Verifying the Cal Status of Reagent Parameters in the Database
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-19
6
Figure 6.8
Reagent Status/Calibration Status Screen
Verifying the Cal Status of Reagent Parameters in the Database
Follow the steps below to verify the calibration status of any reagent loaded in the
database.
E011425S.EPS
Step
Action
1
From the Reagent Status/Calibration Status screen, select
Reagent
Summary [F2]
2
Verify the calibration status under the Cal Status heading for the desired
reagent.
3
For further information about the reagent, complete the following steps.
From the Reagent Summary dialog box, select a button beside a desired
reagent.
THEN
Select the
<Display>
button to view the reagent lot parameters. (Refer to
OR
Select
<Go To>
and type the Chem and Lot number.
4
Select
<OK>
from the Reagent Lot Parameters dialog box when finished
viewing the reagent lot parameters.