Alignment
Straightness of the Sample or Reagent Probe (from side)
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-58
September 2016
7
Use the flex coupling (1) to raise the probe so that the tip of the probe is
lined up with the piece of tape. (Refer to Figure 10.29.)
Figure 10.29
If the tip of the probe is…
then…
centered (1) between the tape
marks,
the probe is straight. (Refer to
Figure 10.30.)
not
centered between the tape
marks,
align probe by gently bending it
to center of the tape marks.
Figure 10.30
8
Reinstall the sample carousel cover.
Step
Action
, continued
(5 of 6)
1
A014078P.EPS
1
A014079P.EPS