Callable Diagnostics
Sample Stir Motor and Reagent Stir Motor
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-50
September 2016
Figure 10.22 Callable Diagnostics Dialog Box
Sample Stir Motor and Reagent Stir Motor
The Sample and Reagent Stir Motor option is used to test or to verify:
• The revolutions per minute of the Sample and Reagent Stir Motors.
• The straightness of the Sample and Reagent Mixer/Paddle.
Follow the steps below to measure motor speed or test paddle straightness.
E014071S.EPS
Step
Action
1
Select
Utilities
from the menu bar.
2
Select
<3> Diagnostics
from the Utilities screen.
OR
Enter
3
in the
Option No.
field.
3
Select
<2> Callable Diagnostics
from the Diagnostics dialog box.
OR
Enter
2
in the
Option No.
field and press
[Enter]
.
(1 of 2)