Configuring the Chemistry Menu
Clearing All 72 Positions
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-5
5
Clearing All 72 Positions
The instrument status must be in
Standby
in order to proceed with the steps below to
clear all positions.
Inserting a Blank Position for Chemistry Insertion
This function is useful for inserting a chemistry into the menu. The subsequent
chemistries will be incremented by one position. After a blank position is inserted, a
chemistry can be defined for that position using the normal procedure. (Refer to
"Defining a Chemistry for a Position" in this section.)
The instrument status must be in
Standby
in order to proceed with the steps below to
insert a blank position.
Step
Action
1
From the Setup screen, select
<1> Chemistry Configuration
.
2
Select
Clear All [F4]
.
3
Select
<OK>
to clear all of the configured chemistries.
OR
Select
<Cancel>
to return to the Chemistry Configuration screen without
clearing all of the configured chemistries.
Step
Action
1
From the Setup screen, select
<1> Chemistry Configuration
.
2
Select a position.
3
Select
Insert Chem [F5]
to insert a blank position and increment the
subsequent chemistries by one position.
4
Define a chemistry for the new blank position as described in "Defining a
Chemistry for a Position," in this section.