Checking Dilution Segment Status and Clearing Dilution Segments
Introduction
System Status/Instrument Commands
IMMAGE 800 Operations Manual A11403
Page 11-2
September 2016
System Status/Instrument Commands
Checking Dilution Segment Status and Clearing Dilution Segments
Introduction
Prior to starting every run, the Dilution Segments status should be checked.
Options
There are two options from the Dilution Segments dialog box:
• The status of the dilution segments can be left unchanged.
• Up to four dilution segments can be cleared. The cleared segments must be replaced
with unused segments on the sample carousel.
Checking Status and Clearing Segments
The instrument status must be in
Standby
in order to proceed with the steps below to
check the status of the dilution segments.
Step
Action
1
Select
Status
from the menu bar.
2
Select
<1> Dilution Segments
3
If dilution segments are...
Then...
to be cleared
Type the segment number(s) in the
Clear Segment(s)
field.
Numbers can be separated by
a comma as a series or by a dash as
a range.
OR
Select the number beside each
appropriate segment
AND
Select
<OK>
.
not
to be cleared
Select
<Cancel>
.