Editing Sample Programs Before Rerunning
Introduction
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-64
September 2016
Editing Sample Programs Before Rerunning
Introduction
After samples are selected to be rerun, the individual sample programs can be edited.
The following sample program information is allowed to be edited:
• chemistries (may be added or deleted)
• STAT check box
• Sample Comment
• Test Replicates
• AGXS testing
• Non-standard dilutions
Reuse Dilution Option
There is an option to reuse the dilution made in the dilution well when a sample was
originally run. (Refer to "Editing Programs Before Rerun," Step 3.) This option can be
useful when sample volume is limited.
Reuse Dilution cannot be used if the necessary dilution segments were cleared or
removed from the sample carousel.
If Reuse Dilution is selected and there is not enough diluted sample available in the
dilution well, the dilution will be remade in a new well.
Note: The default is to remake a dilution.
Clear Chems Option
There is an option to clear all of the chemistries originally selected for a sample.
(Refer to "Editing Programs Before Rerun," Step 3.)
This option can be useful when the original sample program had several chemistries
selected, and only one or two chemistries are to be rerun.
CAUTION
Due to the possibility of sample evaporation over time, the Reuse Dilution feature
should be used with discretion.