Default Setup
Introduction
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-57
5
Default Setup
Introduction
Default Setup is used to:
• define the default sample type for all samples programmed. The sample type can be
changed for individual samples from the Program Sample screen.
• define the default number of replicates to be run for each sample and what statistics
to run on replicates.
• enable or disable statistics to run on replicates.
• define the default setup of the Calibration Halt function.
Setting the Default Sample Type
The instrument status must be in
Standby
in order to proceed with the steps below to
define default sample types.
Step
Action
1
From the Setup screen, select
<11> Default Setup
2
Select options button
<
▼
>
beside
Default Sample Type
.
3
Select the button beside desired default sample type from the choices
below:
•
Serum
•
CSF
•
Plasma
•
Random Urine
•
Timed Urine
The default is serum.
The selection will be shown in the Default Sample Type field.
Select
<OK>
to return to the Setup dialog box and change the sample type.
OR
Select
<Cancel>
to return to the Setup dialog box without changing the
sample type.