Entering Sample Identification
Displaying/Editing a Sample Program
IMMAGE 800 Operations Manual A11403
Sample Programming
September 2016
Page 7-15
7
Displaying/Editing a Sample Program
A previously saved sample program can be displayed by entering either the Rack and
Position or the Sample ID. The program can then be edited.
NOTICE
Every Sample ID must be unique. A Sample ID can only be reused by:
• rerunning the sample. The original and rerun results will be collated.
Example:
A sample is programmed for tests A and B. Test B requires pretreatment
of the sample. The untreated sample can be run first for test A, and the
pretreated sample can then be rerun for Test B only.
OR
• clearing the Sample ID and reprogramming it. The original and
reprogrammed Sample IDs will be treated as two different samples.
Example:
A lab receives two different samples with identical Sample IDs. One of the
samples can be run first and cleared by Sample ID, and the remaining
sample can then be run.