Callable Diagnostics
Sample Stir Motor and Reagent Stir Motor
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-51
10
4
If…
then...
Sample Stir Motor is
desired,
Select
<10> Sample Stir Motor
from
the Callable Diagnostics screen.
OR
Enter
10
in the
Option No.
field and
press
[Enter]
.
Reagent Stir Motor is
desired,
Select
<11> reagent stir motor
from
the callable diagnostics screen.
OR
Enter
11
in the
Option No.
field and
press
[Enter]
.
5
Observe the Mixer/Paddle for sway or wobble during motor exercise.
Replace the Mixer/Paddle if necessary. (Refer to Replacing Parts/User
Servicing, Sample and Reagent Crane Mixer/Paddle in this chapter.)
6
Verify that the Speed of Motor displayed at the end of the motor exercise is
within 6000 to 10000 Revolutions per minute.
Call Beckman Coulter Clinical Support if the "Revolutions per minute" is
out of specifications.
7
Select…
to…
Call. Diag [F1]
Return to the Callable Diagnostics
screen.
Start Motor [F2]
Run the motor continuously.
Stop Motor [F3]
Stop the motor.
Step
Action
, continued
(2 of 2)