Units Setup
Introduction
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-37
5
Units Setup
Introduction
Units can be selected for reporting with the results and displayed throughout the
IMMAGE 800 system for each chemistry.
Selecting Units for Each Chemistry
The instrument status must be in
Standby
in order to proceed with the steps below to
select the desired units for a chemistry.
Step
Action
1
From the Setup screen, select
<7> Units/Non-Standard Dilutions
. (Refer
2
Select the sample type, then select the options button
<
▼
>
beside the
desired chemistry. (Refer to Figure 5.14.) The default units are displayed.
3
Select the number beside the units desired for the selected chemistry and
sample type.
OR
Select
<Cancel>
to return to the Units screen without changing the units.
4
If the units selected are in a different category than the default units, a
Units Conversion dialog box is displayed. (Refer to Figure 5.15 and Table
5.4.) The default units are displayed in the Units Conversion dialog box.
Follow the steps below to enter a conversion factor.
• Type the conversion factor to be multiplied by the default units in the
Conversion factor
field.
• Select
<OK>
to save the conversion factor. OR
• Select
<Cancel>
to return to the
Units
dialog box without saving the
conversion factor.
5
Repeat Steps 2-4 for additional chemistries.