Cartridge-Specific Calibration
Calibration Events
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-25
6
a new reagent lot calibration is
desired,
recalibrate with a fresh cartridge (greater than
predetermined number of tests remaining).
• remove cartridge-specific cartridge
• place a "fresh" cartridge on carousel
• request and recalibrate chemistry
• replace used cartridge on carousel after
successful calibration
specific calibration information
is desired,
select the options button
<
▼
>
beside Cal ID for
the calibrator of the specific chemistry. The
Calibration Information screen will indicate
whether the calibration was cartridge-specific.
If…
Then…
, continued
(2 of 2)