QC Log
Deleting a Result
Quality Control
IMMAGE 800 Operations Manual A11403
Page 9-24
September 2016
Figure 9.7
QC Log by Reagent Lot
Deleting a Result
The instrument status must be in
Standby
in order to proceed with the steps below.
Follow the steps below to delete results from the QC Log cumulative statistics.
E014037S.EPS
Step
Action
1
From the Quality Control screen, select the control desired.
Use the
<Page Up>
/
<Page Down>
buttons to access additional controls.
2
Select
QC Log [F5]
.
3
Enter a Start date and an End date in the date range fields. The default
Start/End date is the current date. Select
<OK>
to choose the default. Use
[Tab]
to toggle between fields.
Select
<OK>
to continue.
OR
Select
<Cancel>
to exit.
4
Select the result by selecting the check box next to the result.
5
Select
Delete Result [F3]
.
6
Enter a maximum of 3 alphanumeric characters in the
Operator ID
field.
Press
[Enter]
.
(1 of 2)