Checking Status and Clearing/Replacing Dilution Segments
Introduction
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-6
September 2016
Checking Status and Clearing/Replacing Dilution Segments
Introduction
Prior to starting every run, the Dilution Segments status should be checked.
Options
There are two options from the Dilution Segments status screen:
• The status of the dilution segments can be left unchanged.
• Up to four dilution segments can be cleared. The cleared segments must be replaced
with unused segments on the sample carousel.
Checking Status and Clearing Segments
The instrument status must be in
Standby
in order to proceed with the steps below to
check the status of the dilution segments.
After Clearing Segments
Always replace the cleared dilution segments with unused segments on the sample
carousel before a run is started.
Step
Action
1
Select
Status
from the menu bar.
2
Select
<1> Dilution Segments
.
3
If dilution segments are...
then...
to be cleared
Type the segment number(s) in the
Clear Segment(s)
field. Numbers can
be separated by a comma as a series
or by a dash as a range.
OR
Select the numbered button beside
each appropriate segment
AND
Select
<OK>
.
not
to be cleared
Select
<Cancel>
.