System Power Off
Introduction
System Power On/Off
IMMAGE 800 Operations Manual A11403
Page 4-4
September 2016
System Power Off
Introduction
It is recommended that the instrument remains powered on and in
Standby
when not in
use.
Powering off the system is recommended for the following situations:
• when parts replacement procedures specify power to be turned off
• when moving the system to a new location
• when lab power goes off (the console is protected by the UPS)
Once the procedure is complete or lab power is restored, power on the system.
Power Off Sequence
The instrument status must be in
Standby
in order to proceed with the steps below to
power off the IMMAGE 800 system.
Emergency Stop
Turn the instrument main power switch off if the stop button on the screen is
unavailable and the instrument must be stopped immediately.
NOTICE
The database may become corrupted if power is turned off before the Power Off
sequence is completed.
Step
Action
1
Check that the floppy diskette drive is empty.
2
Select
Utilities
from the menu bar.
3
Select
<Shutdown>
.
4
When the message
Shutdown Complete
is displayed, turn off the printer,
monitor, CPU (computer), UPS, and instrument.
NOTICE
When an emergency stop or unplanned power loss occurs during a run, and power is
restored within 24 hours, the cuvettes must be washed 1 time before a run can be
started. (Refer to CHAPTER 10,
Utilities
, As-Indicated Maintenance, "Washing
If power is restored after 24 hours, the cuvettes must be replaced. (Refer to CHAPTER 10,
Utilities
, As-Indicated Maintenance, "Replacing Cuvettes.")