Calculations Setup
Selecting Units for Excretion Rate Calculations
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-32
September 2016
Selecting Units for Excretion Rate Calculations
The instrument status must be in
Standby
in order to proceed with the steps below to
change the unit for Excretion Rate calculations.
Restoring Default Excretion Rate Calculations
When the instrument status is in
Standby
, proceed with the steps below to restore all
default excretion rate calculations.
Step
Action
1
From the Setup screen, select
<6> Calculations
.
2
Select
<6> Excretion Rate
.
OR
Enter the number beside Excretion Rate in the
Option
field.
3
Select
View/Edit [F1]
.
4
Select the options button
<
▼
>
beside the
Unit
field.
From the Calculation Unit screen, select the number next to the desired
unit.
OR
Enter the number corresponding to the desired unit
AND
Press
[Enter]
.
5
Select
Calc Summary [F1]
.
OR
Select any icon from the menu bar.
6
From the Warning screen, select
<OK>
to change the unit.
OR
Select
<Cancel>
to exit without changing the unit.
Step
Action
1
From the Setup screen, select
<6> Calculations
2
Select
Select TUrine [F2]
3
Select
<Restore Default>
. All the urine chemistry checkboxes will be
unchecked and the timed urine calculations for those chemistries disabled.
4
Select
<OK>
to return to the Calculations Summary screen.
OR
Select
<Cancel>
to return to the Calculations Summary screen without
changing the timed urine chemistry definition.