Alignment
Sample and Reagent Crane Probe Height/Depth
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-59
10
Sample and Reagent Crane Probe Height/Depth
Follow the steps below to align the Sample or Reagent Crane Probe height/depth.
9
Reinstall the access cover of the Crane Assembly. Position the tubing in
the grooves before tightening the screws.
Step
Action
, continued
(6 of 6)
Step
Action
1
Select
Utilities
from the menu bar.
2
Select
<4> Alignment
.
OR
Enter
4
in the
Option No.
field and press
[Enter]
.
3
If...
then...
aligning the Sample
Probe,
1. Select
<9> Sample Crane Probe
from the Functional Areas dialog
box.
OR
Enter
9
in the
Area No.
field and press
[Enter]
.
2. Select
<1> Sample Crane Probe
from the Alignment dialog box.
OR
Enter
1
in the
Procedure No.
field and
press
[Enter]
.
aligning the Reagent
Probe,
1. Select
<10> Reagent Crane Probe
from the Functional Areas dialog
box.
OR
Enter
10
in the
Area No.
field and press
[Enter]
.
2. Select
<1> Reagent Crane Probe
from the Alignment dialog box.
OR
Enter
1
in the
Procedure No.
field and
press
[Enter]
.
(1 of 2)