Selecting Save/Next
Introduction
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-44
September 2016
Selecting Save/Next
Introduction
When all desired information is programmed for a sample, select
Save/Next [F10]
to
program additional samples.
A message will display if there is not enough information entered to save the sample
program.
If a minimum of a sample identifier (Rack and Position or Sample ID) and a chemistry
is programmed for a sample, the sample program will be automatically saved if the
Program Sample screen is exited without selecting
Save/Next [F10]
.
Automatic Incrementing
If a Rack number and Position number are entered for the current sample, selecting
Save/Next [F10]
will automatically increment to the next position within the rack
according to the following guidelines.
Selecting
Save/Next [F10]
from the last position on a rack will display a blank
Program Sample screen.
Table 7.8
Automatic Incrementing
If the next position on the rack is...
then Save /Next will display...
not programmed,
a blank Program Sample screen with the
current Rack and a new Position.
programmed, but not run,
the corresponding sample program.
complete, incomplete, or
programmed for a rerun or
calibrator,
the next position on the rack that is not
programmed or is programmed but not run.