Cal Options
Introduction
Reagents/Calibration
IMMAGE 800 Operations Manual A11403
Page 6-36
September 2016
Cal Options
Introduction
Cal Options is used to access Calibrator Summary (calibrator lot parameters), Slope
and Offset Adjustments, or Print Last Calibration Results.
Displaying Calibrator Lot Parameters
The instrument status must be in
Standby
in order to follow the steps below to display
calibrator lot parameters.
Figure 6.13
Calibrator Summary Dialog Box
Step
Action
1
Select
Rgts/Cal
from the menu bar.
2
Select
Cal Options [F5]
.
3
Select
<1> Calibrator Summary
4
Select a button beside the Calibrator to be displayed.
THEN
Select the
<Display>
button to view the calibrator lot parameters. (Refer to
OR
Select
<Go To>
and type the Cal and Lot number.
Select
<OK>
from the Calibrator Lot Display dialog box when finished
viewing the calibrator lot parameters.
E010302S.EPS