As-Indicated Maintenance
Decontaminating Carousels and Racks
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-16
September 2016
3
To remove the sample carousel, unscrew the two top and two side screws
(1) of the sample carousel cover.
Figure 10.10
4
Unscrew the sample carousel cap. Remove the sample carousel.
Figure 10.11
5
Prepare a 10% bleach solution (approximately 0.5% final hypochlorite
solution) in deionized water.
6
Immerse carousels or sample racks into a large container or sink filled with
the bleach solution. Allow carousels or racks to soak at room temperature
for 15 to 20 minutes.
7
Remove the carousels or racks and rinse with deionized water or tap water
and let stand until dry.
8
Observe rack labels for bubbling, peeling, or fading. Remove and replace
with a new label if necessary according to instructions in CHAPTER 5,
System Setup
, Placing Bar Code Label on Rack
Step
Action
, continued
(2 of 2)