As-Indicated Maintenance
Replacing Cuvettes
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-17
10
Replacing Cuvettes
Cuvettes are clustered in sectors of three for efficient handling. Handle cuvettes by the
top edges only. Do not remove or touch Reference Cuvette (number 40).
The instrument status must be in
Standby
in order to proceed with the steps below.
Step
Action
1
Verify that the cuvette washer is raised into the cuvette wash station. If the
cuvette washer is lowered into a cuvette, select
<Home>
from the Main
screen to raise the cuvette washer.
2
Remove the two reaction module cover screws (1).
Figure 10.12
3
Slide reaction module cover forward from under cuvette wash station as
shown.
Figure 10.13
(1 of 3)