Errors
Error Table
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-35
10
1620
Mechanical
Devices are
inoperable. Go
to Diagnostics or
Alignment to
correct the
problem.
WARNING: Reagent or
reaction cover may have
been opened during
power on.
1. Check that the reagent cover is on
the instrument.
2. Close reagent cover if open.
3. Select
Utils
.
4. Select
Diagnostics
.
5. Select
Main
.
6. Select
<OK>
to exit the
Diagnostics message.
7. Observe that the probes home.
8. Select
Home
.
9. If the probes home, then proceed
to the next step. If the WARNING
is displayed again,
CALL BECKMAN COULTER
CLINICAL SUPPORT.
10. Wash the cuvettes twice.
1625
CAU Optics Post
Run Check
Failed: NIA
Channel
WARNING:
Repeat all chemistries run on the NIA
Optics.
1626
CAU Optics Post
Run Check
Failed: LPIA
Channel
WARNING:
Repeat all chemistries run on the
LPIA Optics.
1627
Cannot Start
Run-Alignment
Checksum
WARNING:
Reboot the instrument.
1628
Cannot Start
Run-UDS File
Not Loaded
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
1629
File Gets Error:
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
1630
Download Error:
UDS File
Exceeded Limit
of 1024 Lines
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(14 of 21)