Sample and Reagent Crane Probe
Replacing Sample and Reagent Crane Probe
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-77
10
13
Reinstall the crane assembly access cover.
Reinstall the level sense cable receptacle cover.
14
Select any icon from the menu bar to exit Diagnostics.
15
Prime the system. (Refer to Troubleshooting, Daily Maintenance
Procedures, "Priming" in this chapter.)
16
Verify the straightness of the probe. (Refer to Troubleshooting,
Alignment, Straightness of the Sample Probe in this chapter.)
Step
Action
, continued
(6 of 6)