Editing a Control
Accessing a Control to Edit
IMMAGE 800 Operations Manual A11403
Quality Control
September 2016
Page 9-13
9
2
Enter the new Mean value beside the chemistry it corresponds to in the
Assigned Mean
field. Press
[Enter]
.
Enter the "1" SD value beside the chemistry it corresponds to in the
Assigned SD
field. Press
[Enter]
.
A Define/Edit Controls dialog box is displayed when there are new values
entered in the Assigned Mean and/or Assigned SD fields.
Select…
to…
<1> Restore previously defined
mean/SD and continue
exit the Define/Edit dialog box
and restore the previously defined
mean and/or SD value and
continue.
<2> Apply new mean/SD and
continue statistical database
change the mean and/or SD to the
new values. The previous results
and cumulative statistics are
saved.
<3> Delete existing data and
start new statistical database
change the mean and/or SD to the
new values. The previous results
and cumulative statistics are
deleted.
This option is not available if the
control is programmed for a run.
3
Enter the revised constituent code beside the chemistry it corresponds to in
the
Constituent Code
field (4 alphanumeric characters; type leading
zeroes if they are part of the constituent code).
4
Select
Control ID [F2]
.
5
Enter the revised control ID into the
Control ID
field. (Up to 15
alphanumeric characters per field.) At least one Control ID and up to 8
different control IDs may be defined for each control name.
6
Select
<OK>
to save the Control IDs.
OR
Select
<Cancel>
to return to the Define/Edit Controls screen without
saving the Control ID.
7
Select any icon from the menu bar to exit the Define/Edit Controls screen
and save the edited control definition.
Step
Action
, continued
(2 of 2)