Loading Calibrators on the Sample Carousel
Displaying/Printing Cal Load List
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-31
6
Displaying/Printing Cal Load List
After the calibrators are manually assigned, follow the steps below to display or print
the Calibrator Load List and to load calibrators.
Placing Non-bar Coded Calibrators on the Carousel
Follow the steps below to place non-bar coded calibrators on the sample carousel.
Step
Action
1
From the Reagent Status/Calibration Status screen, select
Cal LdList [F6]
.
2
Select
<Print>
to print a Cal Load List.
OR
Select
<OK>
to return to the Reagent Status/Calibration Status screen.
for an example printout).
Step
Action
1
According to the Cal Load List, place a test tube into the appropriate rack
and position.
2
Place the appropriate sample cup in the test tube.
3
Place the appropriate calibrator in the sample cup.
4
Repeat steps 1-3 for any additional calibrators. Continue placing
calibrators into the same sample rack.
5
Open the sample compartment cover.
6
Place the rack(s) containing calibrators on the sample carousel. Rotate the
sample carousel by pressing the advance button on the instrument.
NOTICE
Place racks containing calibrators in position A and B.
Controls and/or patient samples may be placed in any other positions on
the sample carousel. (Refer to CHAPTER 7,
7
Close the sample compartment cover.