Enabling/Disabling Error Display Mode
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-53
5
Enabling/Disabling Error Display Mode
When fatal errors are generated, there is a problem with transmission to or from the
host system. If the error display mode is enabled, these errors will be displayed on the
screen. If the error display mode is disabled, these errors will not be displayed on the
screen but can be found in the Error Log. (Refer to CHAPTER 10,
The instrument status must be in
Standby
in order to proceed with the steps below to
enable or disable error display mode.
Enabling/Disabling Continuous Numbering Mode
The frame is a basic unit of data transmission. When the continuous numbering mode
is checked (enabled), frame numbers from the host will not be reset at "1" with the
next header record.
The instrument status must be in
Standby
in order to proceed with the steps below to
enable or disable the continuous numbering mode.
Step
Action
1
From the Setup screen, select
<10> Host Communications
. (Refer to
2
Select the check box next to
Error Display Mode:
to enable error display
mode.
OR
Deselect the check box to disable error display mode.
Step
Action
1
From the Setup screen, select
<10> Host Communications
. (Refer to
2
Select the check box next to
Continuous Numbering Mode
to enable
continuous numbering.
OR
Deselect the check box to disable Continuous Numbering Mode.