Alignment
Sample and Reagent Crane Probe Height/Depth
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-60
September 2016
Figure 10.31 Alignment Dialog Box
4
Complete Steps 1-3 as described on the screen, selecting
<Continue>
after
performing each step.
OR
Select
<Stop>
to exit. Select
<Close>
to return to the Alignment screen or
select
<Repeat>
to restart the alignment procedure.
5
Select
<Yes>
from the Alignment dialog box to save the alignment data to
a temporary file.
OR
Select
<No>
to return to the Alignment: Reagent Crane Probe screen
without saving the alignment data.
6
Select any icon from the menu bar. (Refer to Figure 10.31.)
Select
<Yes>
from the Alignment dialog box to save the new alignment
data to the database.
[F1] Prev Align
allows the operator to overwrite this
alignment data with the previous alignment data.
OR
Select
<No>
to restore the previous alignment data.
Step
Action
, continued
(2 of 2)
E014080S.EPS