Defining a UDR Chemistry
Beginning a New UDR Protocol Definition
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-79
5
Defining an AGXS Limit
The AGXS Limit is a user-defined value. The AGXS Limit feature will only be
available if the UDR has a status of "calibrated" and uses a "Non-Competitive
Nephelometric" or "Non-Competitive NIPIA" protocol. Entry of this value in the
AGXS Limit field is only available after a UDR multi-point calibration has been
performed and approved.
Entering an AGXS Limit value into the UDR definition will clear the calibration
program of the defined UDR. The AGXS limit is calibration specific; therefore,
recalibrating a UDR using a multi-point calibration will clear the AGXS Limit field
and disable AGXS flagging.
During the sample run, if the calibrated rate for a UDR sample reaction is equal to or
greater than the value entered for the AGXS Limit in the UDR definition, the system
will suppress the results and display ******* in the results column on the results
display and printout instead of a numerical result. AGXS flagging shall not apply
during UDR Rate Mode runs.
9
Select the options button
<
▼
>
beside the
Reagent Expiration Date
field.
10
Enter the reagent expiration date.
NOTICE
The expiration date must not be the current date. Recalibration will be
necessary when the expiration date is changed.
11
Select
<OK>
to enter the expiration date into system.
OR
Select
<Cancel>
to exit the dialog box without entering the date. Continue
to “Defining UDR Sample/Reagent Volumes.”
12
Enter the reagent cartridge serial number in the
Cartridge Serial Number
field. The serial number is found on the bar code label of the UDR
cartridge supplied by Beckman Coulter.
13
Enter the number of tests in the
Tests per Cartridge
field. The maximum
number of tests is 300.
14
If desired, select the
Comments
field and enter vendor information or
other comments. This field is limited to 20 alphanumeric characters.
Step
Action
, continued
(2 of 2)