Programming a STAT
Introduction
IMMAGE 800 Operations Manual A11403
Sample Programming
September 2016
Page 7-43
7
Programming a STAT
Introduction
A sample which requires priority can be programmed as a STAT. The running priority
for samples is:
1. STAT calibration
2. STAT samples/controls
3. Routine calibration
4. Routine samples/controls
A STAT sample is run before any routine patient or control sample on the carousel,
regardless of the Rack and Position numbers.
A STAT calibration is run before a STAT sample. A STAT calibration must be
programmed when the calibration is requested.
If multiple STAT samples are programmed, they will be run in the order they are
placed around the sample carousel, in a counter-clockwise direction.
Programming a STAT Sample
Follow the steps below to program a STAT sample.
Additional Information
Refer to CHAPTER 6,
Reagents/Calibration
, Requesting and Canceling Calibration,
for more information on programming a STAT calibration.
Step
Action
1
Select
Samples
from the menu bar.
2
Select the
STAT
check box. This can be done at any time while
programming a sample.