Calibrating a UDR Chemistry
Programming a UDR Calibration
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-96
September 2016
Figure 5.41
UDR Request Cal Dialog Box
Figure 5.42
UDR Multi-point Cal Assign Screen
7
Select
Save [F9]
or a menu bar icon to save the calibration program for the
displayed chemistry.
OR
Select
Cancel [F10]
to exit the screen without saving any calibration
programs.
8
Repeat Steps 4-7 to program additional racks for calibration of other UDR
chemistries.
Step
Action
, continued
(2 of 2)
E014051S.EPS
E014052S.EPS