Configuring Antigen Excess Testing
Introduction
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-42
September 2016
Configuring Antigen Excess Testing
Introduction
Antigen excess (AGXS) testing can be enabled or disabled for each appropriate
chemistry configured on the chemistry menu.
The default for AGXS testing is enabled for all the appropriate chemistries.
If AGXS testing is enabled, AGXS testing is always performed for the associated
chemistry. If AGXS testing is disabled, AGXS testing will not be performed for the
associated chemistry. AGXS can be enabled or disabled for an individual sample in
Sample Programming.
How to Configure AGXS Testing
The instrument status must be in
Standby
in order to proceed with the steps below to
configure AGXS testing.
Figure 5.17
Antigen Excess Screen
Step
Action
1
From the Setup screen, select
<8> Antigen Excess
. A multipage list of
chemistries will be displayed. (Refer to Figure 5.17.)
2
A checkmark in the box indicates AGXS is enabled for that chemistry.
Select the check box to remove the check and disable the AGXS testing.