Instrument Specifications
Placement
System Description
IMMAGE 800 Operations Manual A11403
Page 2-46
September 2016
System Specifications and Characteristics
Instrument Specifications
Placement
The surface on which the unit rests must be free of vibration and must be level, 1° or
<0.75 inch (1.9 cm) slope across the length and the width of the instrument. Do not
place instrument in direct sunlight or drafts or near a heating or cooling duct.
Clearance
Sides - 6 inches (15.2 cm) minimum
Back - None
Front - 3 inches (7.6 cm) minimum
Top - 4 inches (10.1 cm) from top of instrument
Dimensions (Excluding Wash and Waste Bottles)
Height = 30 inches (76.2 cm)
Depth = 25.5 inches (64.8 cm)
Length = 43.5 inches (110.5 cm)
Weight
250 lb. (120 kg)
Power Requirements
Supplemental Specifications
Operating Range
100-120VAC, 200-240VAC Single Phase
Frequency Range
50/60 Hz
Transient Suppression
Recommended
BTU Generated
2,900 BTU/hour
Electrical Outlet
Grounded per Local Code
Surge Protector
Recommended
Current
8.0 Amps (normal) 12 Amps surge
Environment
Indoor use only
Installation Category
II
Pollution Degree
2
EN55011
Meets class A
Maximum Sound Pressure per IEC 61010-1
Normal: 82 dBA