Sample and Reagent Crane Probe
Replacing Sample and Reagent Crane Probe
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-76
September 2016
9
Remove the set screw on the new probe. (Refer to Figure 10.43.) Install
the new probe into the crane assembly.
Figure 10.43
NOTICE
Align the hole of the probe with the access hole (1) in the RF housing.
(Refer to Figure 10.44.)
Figure 10.44
10
Secure the probe by reinstalling the set screw.
11
Connect the tubing from the top of the probe assembly.
12
Connect the level sense cable to the receptacle.