Checking Status and Clearing/Replacing Dilution Segments
Replacing Dilution Segments
IMMAGE 800 Operations Manual A11403
Sample Programming
September 2016
Page 7-7
7
Replacing Dilution Segments
The instrument status must be in
Standby
in order to proceed with the steps below to
replace dilution segments.
Figure 7.1
Replacing a Dilution Segment
Additional information
System Status/Instrument Commands
for more information on
Dilution Segments status.
Step
Action
1
Lift the cover of the sample carousel.
2
Rotate the sample carousel by pressing the advance button in order to
access the desired dilution segment, if necessary.
3
Lift the segment to be replaced off of the sample carousel. (Refer to
Figure 7.1.)
4
Dispose of the used segment in a manner appropriate for biohazardous
materials.
5
Place an unused segment in the empty position on the sample carousel.
6
Repeat Steps 2-5 until all desired segments are replaced.
1. Dilution Segment
A011427P.EPS