Calculations Setup
Selecting Urine Chemistries for Excretion Rate Calculations
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-31
5
Selecting Urine Chemistries for Excretion Rate Calculations
After the excretion rate calculation is enabled, the urine chemistries for the calculation
must be enabled. Once enabled, the excretion rate calculation will utilize the hours
and volume entered for the urine sample.
When the instrument status is in
Standby
, proceed with the steps below to enable urine
chemistries for excretion rate calculations.
Figure 5.12
Define Timed Urine Chemistries Dialog Box
Step
Action
1
From the Setup screen, select
<6> Calculations
2
Select
Select TUrine [F2]
3
Select the check box beside the urine chemistry desired for the excretion
rate calculation. The default is disabled (unchecked).
4
Select
<OK>
to return to the Calculations Summary screen.
OR
Select
<Cancel>
to return to the Calculations Summary screen without
changing the urine chemistry definition.
E010307S.EPS