Errors
Error Table
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-24
September 2016
E24
P24VD PS OR
P24VD Power Supply
Out of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E25
P24VE PS OR
P24VE Power Supply
Out of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E26
P24VH PS
P24VH Power Supply
Out of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E27
P24VS PS
P24VS Power Supply
Out of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E28
P24VP PS
P24VP Power Supply
Out of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E29
Hi Press
High Pressure Out of
Range
With system running samples:
1. Check Status Monitor.
2. Verify high pressure is in range.
E30
Lo Press
Low Pressure Out of
Range
With system running samples:
1. Check Status Monitor.
2. Verify low pressure is in range.
E31
Vac
Vacuum Out of Range
With system running samples:
1. Check Status Monitor.
2. Verify vacuum is in range.
E32
RGT A LVL
Level Sense Failure,
Reagent Compartment A
1. Check and remove bubbles from
Reagent Compartment A.
2. Rerun sample.
E33
RGT B LVL
Level Sense Failure,
Reagent Compartment B
1. Check and remove bubbles from
Reagent Compartment B.
2. Rerun sample.
E34
Buff LVL
Level Sense Failure on
Buffer
1. Check and remove bubbles from
buffer.
2. Observe that reagent probe enters
buffer.
3. Rerun sample.
E35
Dil LVL
Level Sense Failure on
Diluent
1. Check and remove bubbles from
diluent.
2. Rerun sample.
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(3 of 21)