Sample and Reagent Crane Mixer/Paddle
Replacing Sample and Reagent Crane Mixer/Paddle
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-71
10
Sample and Reagent Crane Mixer/Paddle
Replacing Sample and Reagent Crane Mixer/Paddle
Follow the steps below to replace the Sample or Reagent Crane Mixer/Paddle.
Step
Action
1
Rotate the Crane Assembly toward the sample carousel.
2
Remove the Crane Mixer/Paddle (1) by gently pulling it down and out from
the crane assembly. (Refer to Figure 10.35)
Figure 10.35
3
Install the new mixer/paddle, making sure that it is pushed all the way up
and properly seated.
4
Verify the straightness of the mixer/paddle. (Refer to Troubleshooting,
Callable Diagnostics, "Sample Stir Motor and Reagent Stir Motor" in this
chapter.)
1
A014083P.EPS