Alignment
Straightness of the Sample or Reagent Probe (from side)
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-57
10
5
Use the flex coupling to lower the probe so that the upper, large diameter
portion of the probe falls just inside the compartment and is centered on
the tape.
6
From the side of the instrument, make marks (1) and (2) on the piece of
tape where the sides of the probe line up. (Refer to Figure 10.28.)
Figure 10.28
Step
Action
, continued
(4 of 6)
1
2
A014077P.EPS