Default Setup
Definition of Replicates
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-58
September 2016
Figure 5.22
Default Setup Dialog Box
Definition of Replicates
System Replicates – The number of times each test will be performed on each sample.
Setting System Replicates
The instrument status must be in
Standby
in order to proceed with the steps below to
set the desired number of system replicates.
E014092S.EPS
Step
Action
1
From the Setup screen, select
<11> Default Setup
2
Type the desired number of system replicates (1 to 9) in the System
Replicates
field.
3
Select
<OK>
to return to the Chemistry Configuration screen.
OR
Select
<Cancel>
to return to the Chemistry Configuration screen without
changing the statistics setup.